T stability of focal plane
ADR, thus we should not assume same stability as Planck.
The concern is that the real numbers we use
Loop gain = 1000 from thesis, very high, Roger think L~10 is more reasonable.
Alpha is also too large by a factor of 10 or more.
Toki agrees with Roger. Alpha is about ~100.
For LiteBird, super relaxed, requirement For T stability is very low.
Shaul: LiteBird no ADR, we don’t know what it should be for continuous ADR.
Roger: 1%Single ADR would heat the focal plane.
Shaul: timescale matters. We can calibrate on dipole. If the fluctuation is fast, then it’s problematic. On spin-spin basis (~1min), we would have dipole calibration. Timescale is unknown for CADR.
LiteBird: single shot, not ADR. ADR had been discussed. Toki is going to dig out more information.
Shaul: we will get information from Toki aobut LiteBird.
Roger: by next week, If we have a figure from Toki on requirement, we can have a number, we can put a reqirement on the fridge. Temperature constrained within some range.We are looking for “changes in detector response over some timescales”.
Roger: if you cycle ADR, before next cycle, you can do calibration.
Amy: on one hand, amount of requirement on changing responsivity and time scales; use calculation to infer Bath stability and timescale; we don’t need to go into too much details at this stage. More importantly, Decal Panel, risky or not.
Shaul: have no idea wether this is an issue or not; we don’t know ADR, fluctuation and timescale. Goddard may care cost, then they give us amplitude and timescale. Shaul will follow up.
“Common modes”, If the two polarization go down and up together.
Shaul will write to Tom, and with information Toki will proved, we will see what we need.
Kris: since this is a study, other techonologies could be mentioned.
Shaul: we will say there are emerging technolgy that may end being competative.