====== Wet Chemistry ====== - Description/Name: Wet Chemistry Lab - Planned use (be specific): Material Assay and Samples Preparation for Low Background Counting - Primary function: the 4850-ft level - Temporary and permanent storage: Chemical Cabinet - Sample and detector handling: Nitric acid, clean solution - Jigs, fixtures, scaffolds, etc. necessary : two tables - E & O activities: Training studnets - Occupancy: - Steady: 1-2 person - Peak: 10 persons - Adjacencies (list/distance): 1-2 meters - Space size (gsf and/or WxLxH): 6 m x 5.5 m x 3 m (33 m^2) - Experiment Systems: lights and clean towels - Shielding: No - Floor load (PSF): Round up for safety ~ 1.5 tons/ft2. - Material handling (cranes, etc.) Movable crane/floor rail for heavy item movement - Finishes (walls/floor/ceiling): Class 10,000 clean standard - Casework: - Storage: Chemical stuff - Furniture/Fixtures/Equipment (non-experiment): Tables - Mechanical: - Environmental conditions: Radon reduced air, temperature and humility - Clean class: 10,000 - Ventilation: 100 cfm - Exhaust: 100 cfm - Plumbing (domestic, waste, systems): water supply, heating and cooling - Fire Suppression: - Building Management System: rest room for changing PPE - Fume hoods: 1 (ea 4x5ft^2=20ft^2: 20ft^2) - hoods, storage cabinets, etching tanks and rinse tanks for cleaning lead, glove box - Chilled water: Yes - Other: - Electrical: - Power (clean and dirty): 10 kw fro air condition - Lighting: room lights - Low Voltage/Data/Communication: - Fire Detection/Alarm: - Other: - Acoustical: - Vibration Isolation: - EMI/RFI Isolation: - Hazardous materials: waste disposal, hazardous chemical, lead contaminated ppe disposal, OSHA compliant layout - Special Consideration (expansion, etc.):