Gamma Counting Stations
Description/Name: Gamma-Ray Counting Station at the 4850-ft level
Planned use (be specific):
Primary function: Materials screening
Temporary and permanent storage: *Separated sample storage rooms for screening and for keeping
Sample and detector handling: Clean room with class 10,000, liquid nitrogen, sample preparation including assay materials
Jigs, fixtures, scaffolds, etc. necessary: Jigs for sample preparation; lifting system for open and close the counting stations. Lab fixtures: light-duty; power, lights, laboratory casework, communications (phone/computer), standard wall coverings, uninterruptible power, 15 kW of power (UPS), air monitoring for certain gases
E & O activities: daily (emergency 24/7); Utilities: air, water, heat/cooling, light, communications, commercial drains, chemical drain/disposal path, liquid nitrogen, humidity control; Radon free air; Temperature; humidity control.
Occupancy:
Steady: Two people
Peak: Ten people
Adjacencies (list/distance):
Space size (gsf and/or WxLxH):20 m x 50 m x 4 m
Experiment Systems: Germanium low background detectors and DAQ system
Shielding: Very stout floor for shield; Copper, lead, poly bricks, water walls
Floor load (PSF): (Pb ~12 g/cm3) a 1-m high shield would be 1,200 g/cm2 or about 2400 lbs/ft2. Round up for safety ~ 1.5 tons/ft2.
Material handling (cranes, etc.) Movable crane/floor rail for heavy item movement
Finishes (walls/floor/ceiling):) Class 10,000 clean standard
Casework:
Storage: purge cabinets; LN2
Furniture/Fixtures/Equipment (non-experiment): chemical storage cabinets, storage cabinets, etching tanks and rinse tanks for cleaning lead, a)Glove box (ea 3x3ft^2=9ft^2: 9ft^2); b)Assembly Table: 4 (ea 5 x 5 ft^2 = 25 ft^2: 100 ft^2) (with test equipments, can be combined to form single larger assembly support tables); c) Chairs: 4 (ea 3 x 3 ft^2 = 9 ft^2: 32 ft^2); d)Electronics Racks: 1 (3 x 2 ft^2 = 6 ft^2: 6 ft^2); e)Parts Storage Cabinets: 4 (4 x 2 ft^2 = 8 ft^2: 64ft^2); Chemical storage cabinets 4 (4 x 2 ft^2 = 8 ft^2: 64ft^2).
Mechanical:
Environmental conditions: Radon reduced air, temperature and humility control
Clean class: 10,000
Ventilation: ~100 cfm
Exhaust: ~ 100 cfm
Plumbing (domestic, waste, systems): water, heating/cooling
Fire Suppression:
Building Management System: Preparation room for changing PPE and keeping the note books
Fume hoods: 1 (ea 4x5ft^2=20ft^2: 20ft^2) - hoods, storage cabinets, etching tanks and rinse tanks for cleaning lead, glove box
Chilled water
Other:
Electrical:
Power (clean and dirty): clean 15 kw
Lighting: room lights
Low Voltage/Data/Communication:
Fire Detection/Alarm:
Other:
Acoustical:
Vibration Isolation:
EMI/RFI Isolation:
Hazardous materials: General waste disposal: 1 hazardous chemical disposal: 1 lead contaminated ppe disposal: OSHA compliant layout
Special Consideration (expansion, etc.):