| - | ====== Surface Infrastructure ====== | ||
| - | |||
| - | |||
| - | - Description/ | ||
| - | - Planned use (be specific): Low Background Counting with Neutron Activation and ICPMS | ||
| - | - Primary function: Gamma-ray counting at the surface | ||
| - | - Temporary and permanent storage: Sample storage and chemical cabinet | ||
| - | - Sample and detector handling: Material assay | ||
| - | - Jigs, fixtures, scaffolds, etc. necessary: light-duty; power, lights, laboratory casework, communications (phone/ | ||
| - | - E & O activities: | ||
| - | - Occupancy: | ||
| - | - Steady: 1-2 persons | ||
| - | - Peak: 4-6 persons | ||
| - | - Adjacencies (list/ | ||
| - | - Space size (gsf and/or WxLxH): 4m x 4m x 3m | ||
| - | - Experiment Systems: Germanium detectors, DAQ, muon veto, and ICMPS | ||
| - | - Shielding: Lead, copper | ||
| - | - Floor load (PSF): (Pb ~12 g/cm3) a 1-m high shield would be 1,200 g/cm2 or about 2400 lbs/ft2. Round up for safety ~ 1.5 tons/ft2. | ||
| - | - Material handling (cranes, etc.): Movable crane/floor rail for heavy item movement | ||
| - | - Finishes (walls/ | ||
| - | - Casework: | ||
| - | - Storage: purge cabinets; LN2 | ||
| - | - Furniture/ | ||
| - | - Mechanical: | ||
| - | - Environmental conditions: Radon reduced air, temperature and humility | ||
| - | - Clean class: 10,000 | ||
| - | - Ventilation: | ||
| - | - Exhaust: 100 cfm | ||
| - | - Plumbing (domestic, waste, systems): water, heating/ | ||
| - | - Fire Suppression: | ||
| - | - Building Management System: rest room for changing PPE | ||
| - | - Fume hoods: 1 (ea 4x5ft^2=20ft^2: | ||
| - | - Chilled water | ||
| - | - Other: | ||
| - | - Electrical: | ||
| - | - Power (clean and dirty): 15 KW | ||
| - | - Lighting: general | ||
| - | - Low Voltage/ | ||
| - | - Fire Detection/ | ||
| - | - Other: | ||
| - | - Acoustical: | ||
| - | - Vibration Isolation: | ||
| - | - EMI/RFI Isolation: | ||
| - | - Hazardous materials: waste streams: | ||
| - | - Special Consideration (expansion, etc.): | ||