-====== Surface Infrastructure ====== 
- 
- 
-  - Description/Name: Surface Infrastructure 
-  - Planned use (be specific): Low Background Counting with Neutron Activation and ICPMS 
-     - Primary function: Gamma-ray counting at the surface 
-     - Temporary and permanent storage: Sample storage and chemical cabinet 
-     - Sample and detector handling: Material assay  
-     - Jigs, fixtures, scaffolds, etc. necessary: Standard lab requirements 
-     - E & O activities:  : Training students and accommodate outreach actuivities 
-  - Occupancy:  
-     - Steady: 1-2 persons 
-     - Peak: 4-6 persons 
-  - Adjacencies (list/distance): 
-  - Space size (gsf and/or WxLxH): 4m x 4m x 3m 
-  - Experiment Systems: Germanium detectors, DAQ, muon veto, and ICMPS 
-  - Shielding: Lead, copper 
-  - Floor load (PSF): (Pb ~12 g/cm3) a 1-m high shield would be 1,200 g/cm2 or about 2400 lbs/ft2. Round up for safety ~ 1.5 tons/ft2. 
-  - Material handling (cranes, etc.): Movable crane/floor rail for heavy item movement 
-  - Finishes (walls/floor/ceiling): Class 10,000 clean standard 
-  - Casework: 
-  - Storage: purge cabinets; LN2 
-  - Furniture/Fixtures/Equipment (non-experiment):hoods, chemical storage cabinets, storage cabinets, etching tanks and rinse tanks for cleaning lead, a)Glove box (ea 3x3ft^2=9ft^2: 9ft^2), b)Assembly Table: 4 (ea 5 x 5 ft^2 = 25 ft^2: 100 ft^2) (with test equipments, can be combined to form single larger assembly support tables), c)Chairs: 4 (ea 3 x 3 ft^2 = 9 ft^2: 32 ft^2), d)Electronics Racks: 1 (3 x 2 ft^2 = 6 ft^2: 6 ft^2), e)Parts Storage Cabinets: 4 (4 x 2 ft^2 = 8 ft^2: 64ft^2), f)Chemical storage cabinets 4 (4 x 2 ft^2 = 8 ft^2: 64ft^2) 
-  - Mechanical: 
-     - Environmental conditions: Radon reduced air, temperature and humility  
-     - Clean class: 10,000 
-     - Ventilation: 100 cfm 
-     - Exhaust: 100 cfm 
-     - Plumbing (domestic, waste, systems): water, heating/cooling 
-     - Fire Suppression: 
-     - Building Management System: rest room for changing PPE 
-     - Fume hoods: 1 (ea 4x5ft^2=20ft^2: 20ft^2) - hoods, storage cabinets, etching tanks and rinse tanks for cleaning lead, glove box 
-     - Chilled water 
-     - Other:  
-  - Electrical: 
-     - Power (clean and dirty): 15 KW 
-     - Lighting: general 
-     - Low Voltage/Data/Communication:  
-     - Fire Detection/Alarm: 
-     - Other:  
-  - Acoustical: 
-  - Vibration Isolation: 
-  - EMI/RFI Isolation: 
-  - Hazardous materials: waste streams:  General waste disposal: 1  hazardous chemical disposal: 1 lead contaminated ppe disposal: OSHA compliant layout 
-  - Special Consideration (expansion, etc.): 
  