 Questions to be answered:  Questions to be answered: 
-  * range of sensitivities + 
 +  
 +* range of sensitivities  
 +    i. Common materials screening: ~10-11 g/g for 40K, 232Th, 238U. 
 +    ii. Very low background: ~10-12 g/g for 40K, 232Th, 238U. 
 +    iii.Ultra-low background in electroforming copper: ~10-13g/ for 232Th, 238U.
   * # detectors   * # detectors
 +    i. 10 gamma-ray stations for common materials screening.
 +    ii. 4 gamma-ray stations for very low background materials.
 +    iii.2 gamma-ray stations for ultra-low background counting.
   * distribution by depth   * distribution by depth
 +    i. 300-ft or 800-ft levels for common materials screening.
 +    ii. 4850-ft for very low-background and ultra-low background counting.
   * staffing and expertise   * staffing and expertise
 +    i. Two staff numbers are needed for operating the facility.
 +    ii. Expertise:  Much expertise can be called from different institutions. 
  
 Software: Reyco Henning Software: Reyco Henning