Line 45: | Line 45: | ||
__Solenoid Valves__- These are the six toggle switches located to the right of the pump controls. In auto mode, they are monitors that display the states of the respective valves ((There is no feedback from the solenoid valves. So the valve states as indicated here are merely the control signals the computer is sending to the valves. If everything is working properly the actual valve state should mirror the control signal and this detail can be ignored.)). Up indicates open and down indicates closed. In override, these six switches become controls allowing the user to manually open and close the valves. | __Solenoid Valves__- These are the six toggle switches located to the right of the pump controls. In auto mode, they are monitors that display the states of the respective valves ((There is no feedback from the solenoid valves. So the valve states as indicated here are merely the control signals the computer is sending to the valves. If everything is working properly the actual valve state should mirror the control signal and this detail can be ignored.)). Up indicates open and down indicates closed. In override, these six switches become controls allowing the user to manually open and close the valves. | ||
- | __Filter Display__- When this button is pressed, the front panel monitors will be low pass filtered (smoothed). The RC time constant can be changed by clicking the " | + | __Filter Display__- When this button is pressed, the front panel monitors will be low pass filtered (smoothed). The equivalent |
__Override Switches__- These switches are intended to be used in auto mode when a specific safety prevents the pump from running continuously. See the section below for more information regarding the safeties. | __Override Switches__- These switches are intended to be used in auto mode when a specific safety prevents the pump from running continuously. See the section below for more information regarding the safeties. | ||
Line 101: | Line 101: | ||
The limits can be changed while the controller is running by clicking on “Limits and Setpoints” on the front panel. A new window will appear and all pump stop criteria is adjustable from the "Pump Stop Criteria" | The limits can be changed while the controller is running by clicking on “Limits and Setpoints” on the front panel. A new window will appear and all pump stop criteria is adjustable from the "Pump Stop Criteria" | ||
- | In certain situations, one or more safeties may need to be temporarily bypassed. Such is common if the pump has been left off for an extended period of time and an appreciable amount of O2 has seeped into the system. In this case the KR and O2 readings may be outside their limits. A safety can always be bypassed by setting the corresponding limit arbitrarily large or small. For example, if the KR is malfunctioning and reading very low, the KR Lower Limit can be set to 0 or some negative value that it will never reach. Alternatively, | + | In certain situations, one or more safeties may need to be temporarily bypassed. Such is common if the pump has been left off for an extended period of time and an appreciable amount of O2 has seeped into the system |
- | Unique to the KR override is an additional | + | Unique to the KR override is an auto reset option. When this switch is turned ON, the KR override will stay on only until the reading falls within the range set by the limits or a timeout period |
===Absolute Limits=== | ===Absolute Limits=== | ||
Line 135: | Line 135: | ||
In any case, one should avoid letting the pump run unattended with any of the overrides switched to on. As an exception, the KR override can be left on provided its auto reset switch is also on. | In any case, one should avoid letting the pump run unattended with any of the overrides switched to on. As an exception, the KR override can be left on provided its auto reset switch is also on. | ||
- | The program itself should be left running nearly 24/7 regardless of whether the pump is on. Even when the pump is off, the program opens and closes the top-up valve to maintain correct pressure in the return manifold. Of course its ok to temporarily quit with dignity and close the program to do computer maintenance such as installing OS updates. However, avoid leaving the program off more than a few hours at a time. | + | The program itself should be left running nearly 24/7 regardless of whether the pump is on. Even when the pump is off, the program opens and closes the top-up valve to maintain correct pressure in the return manifold. Of course its ok to temporarily quit with dignity and close the program to do computer maintenance such as installing OS updates. However, avoid leaving the program off more than a few hours at a time. Any event in which the program is not running for more than a couple hours at a time, the data analysis team should be notified that pressure regulation was temporarily lost, usually by submitting an ELOG entry. |
- | A known issue with system is the tendency for sensors to hiccup. The voltage output of a sensor as read by the computer will sometime abruptly jump well outside the limit set for the sensor. In these cases, the actual condition at the sensor is ok and there is nothing wrong with the gas system itself. The duration of these hiccups are usually short enough such that they will not pass through the software' | + | When the pump shuts down it's ok to try restarting it as the cause of shutdown |
- | + | ||
- | In any case when the pump shuts down it' | + | |
If turning on the pump CAUSES any of the pressures to rise outside their limits, the pump should be turned off immediately. This most likely indicates a mechanical blockage downstream of the pump, such as a jammed closed recirculation valve. The over-pressurization caused by running the pump under these conditions could cause a catastrophic failure and hence a huge leak. | If turning on the pump CAUSES any of the pressures to rise outside their limits, the pump should be turned off immediately. This most likely indicates a mechanical blockage downstream of the pump, such as a jammed closed recirculation valve. The over-pressurization caused by running the pump under these conditions could cause a catastrophic failure and hence a huge leak. | ||
Line 220: | Line 218: | ||
A backup of the Gas Rack folder containing the necessary controller VIs, configuration files and the executable: | A backup of the Gas Rack folder containing the necessary controller VIs, configuration files and the executable: | ||
- | {{: | + | {{: |
- | The main VI (gas_rack_controller.vi) is in the folder " | + | The main VI (gas_rack_controller.vi) is in the folder " |
+ | |||
+ | The LabVIEW code was written in LabVIEW 2012. If using the .exe instead of the VI, LabVIEW itself does not need to be installed but you must have the LabVIEW run-time engine (version 2012f3) and DAQmx 9.4.0 installed. | ||
+ | |||
+ | Also included is the firmware code to program the CPLD located on the control board located inside the blue panel: | ||
+ | |||
+ | {{: | ||
For additional information on the gas re-circulation system contact Jeff Gunderson at gund0328@umn.edu or 612-669-7962. | For additional information on the gas re-circulation system contact Jeff Gunderson at gund0328@umn.edu or 612-669-7962. | ||
Line 248: | Line 252: | ||
{{: | {{: | ||
- | ===Programming control=== | + | The CPLD configuration file is posted at the end of this manual along with the software written in LabVIEW. |
- | + | ||
- | [Joe is currently in the process of finding a copy of the CPLD programming to post right here. Upon repetitive failure to find the copy, this subsection will be omitted.] | + | |
===Software Control=== | ===Software Control=== | ||
Line 354: | Line 356: | ||
====References and Additional Information ==== | ====References and Additional Information ==== | ||
- | Here is a backup | + | A backup |
+ | |||
+ | {{: | ||
+ | |||
+ | |||
+ | The main VI (mixer_controllerYYYY-MM-DD.vi) is included along with the necessary SubVIs. A copy of the pre-compiled executable along with a few helper files is alongside these files in another folder | ||
+ | |||
+ | The LabVIEW code was lasted edited in LabVIEW 2012. If using the .exe instead of the VI, LabVIEW itself does not need to be installed but you must have the LabVIEW run-time engine (version 2012f3) and DAQmx 9.5.1 installed. | ||
+ | |||
+ | The CPLD firmware is also included: | ||
+ | |||
+ | {{: | ||
+ | |||
+ | There is an up-to-date version of the .jed configuration file. The source code included is not the most recent version but is the only version I have been able to find. | ||
+ | |||
+ | |||
- | {{: | ||
- | For additional information on the gas mixing system try Joe Pastika at past0035@umn.edu. |