Surface Infrastructure
Description/Name: Surface Infrastructure
Planned use (be specific): Low Background Counting with Neutron Activation and ICPMS
Primary function: Gamma-ray counting at the surface
Temporary and permanent storage: Sample storage and chemical cabinet
Sample and detector handling: Material assay
Jigs, fixtures, scaffolds, etc. necessary: Standard lab requirements
E & O activities: : Training students and accommodate outreach actuivities
Occupancy:
Steady: 1-2 persons
Peak: 4-6 persons
Adjacencies (list/distance):
Space size (gsf and/or WxLxH): 4m x 4m x 3m
Experiment Systems: Germanium detectors, DAQ, muon veto, and ICMPS
Shielding: Lead, copper
Floor load (PSF): (Pb ~12 g/cm3) a 1-m high shield would be 1,200 g/cm2 or about 2400 lbs/ft2. Round up for safety ~ 1.5 tons/ft2.
Material handling (cranes, etc.): Movable crane/floor rail for heavy item movement
Finishes (walls/floor/ceiling): Class 10,000 clean standard
Casework:
Storage: purge cabinets; LN2
Furniture/Fixtures/Equipment (non-experiment):hoods, chemical storage cabinets, storage cabinets, etching tanks and rinse tanks for cleaning lead, a)Glove box (ea 3x3ft^2=9ft^2: 9ft^2), b)Assembly Table: 4 (ea 5 x 5 ft^2 = 25 ft^2: 100 ft^2) (with test equipments, can be combined to form single larger assembly support tables), c)Chairs: 4 (ea 3 x 3 ft^2 = 9 ft^2: 32 ft^2), d)Electronics Racks: 1 (3 x 2 ft^2 = 6 ft^2: 6 ft^2), e)Parts Storage Cabinets: 4 (4 x 2 ft^2 = 8 ft^2: 64ft^2), f)Chemical storage cabinets 4 (4 x 2 ft^2 = 8 ft^2: 64ft^2)
Mechanical:
Environmental conditions: Radon reduced air, temperature and humility
Clean class: 10,000
Ventilation: 100 cfm
Exhaust: 100 cfm
Plumbing (domestic, waste, systems): water, heating/cooling
Fire Suppression:
Building Management System: rest room for changing PPE
Fume hoods: 1 (ea 4x5ft^2=20ft^2: 20ft^2) - hoods, storage cabinets, etching tanks and rinse tanks for cleaning lead, glove box
Chilled water
Other:
Electrical:
Power (clean and dirty): 15 KW
Lighting: general
Low Voltage/Data/Communication:
Fire Detection/Alarm:
Other:
Acoustical:
Vibration Isolation:
EMI/RFI Isolation:
Hazardous materials: waste streams: General waste disposal: 1 hazardous chemical disposal: 1 lead contaminated ppe disposal: OSHA compliant layout
Special Consideration (expansion, etc.):